Heat meter RTD temperature probes are used for temperature measurement in closed pipeline systems. Due to the installation in immersion sleeves with fitting tolerance, the system no longer needs to be emptied during regular replacement once the certification period has elapsed.
general technical data | |
---|---|
Minimum immersion depth | 25 mm |
Areas of application | Heat and cold meters |
Insertion length | 85 to 400 mm |
Process connection | Thread G 1/2, stainless steel |
Operating temperature | 0 to 180 °C |
Temperature difference | 3 to 180 K |
Terminal head | Type J |
Features | Paired according to European MID directive and/or German law on weights and measures (MessEG) including declaration of conformity (conformity and additional metrology marking). Production certified according to module D of the MID and German ordinance on weights and measures (MessEV). |
Protection fitting | Stainless steel 6 mm |
Approvals | MID and domestic type examination certificates for temperature probes for heat meters, cold meters, and combined cold/heat meters; fulfills the requirements of DIN EN 1434, AGFW FW 202 FW 212, TR K8, and TR K11 |
Temperature sensor | Pt100, Pt500, Pt1000 |
Option | Two-wire or four-wire circuit, terminal head, type B |