The membrane-covered, amperometric sensor works on the three-electrode principle; it is used for the online measurement of hypobromous acid (HOBr). The electronic components integrated in the sensors provide a temperature-compensated current signal (4 to 20 mA) or a Modbus RTU output signal.
general technical data | |
---|---|
Measuring range | 0.05 to 20 mg/l (ppm) |
Measuring principle | Membrane-covered, amperometric, potentiostatic three-electrode system with integrated electronic components |
Inflow speed | Approx. 15 cm/s |
Response time t90 | Approx. 2 minutes |
Operating temperature | 0 to 45 °C |
Settling time | Approx. 2 hrs for initial startup |
pH value area of application | pH 6.5 to pH 9.5 |
Application areas | Disinfection of drinking water, swimming pool water, service water, process water, and seawater |