Sensor for chlorine dioxide and ozoneJUMO tecLine ClO2 and O3

These membrane-covered amperometric sensors measure the concentration of chlorine dioxide or ozone in aqueous solutions (e.g. drinking water, swimming pool water, service water, process water, and cooling water). The sensors can be connected directly to a suitable indicating or control device.

Integrated temperature compensation

Easy calibration

Two-electrode principle

Sensor for chlorine dioxide and ozone JUMO tecLine ClO2 and O3
JUMO tecLine ClO2 and O3 - Sensor for chlorine dioxide and ozone

Technical data

general technical data
Measuring range Chlorine dioxide: max. 0.05 to 10 mg/l (analog); max. 0.05 to 20 mg/l (digital); ozone: max. 0.05 to 20 mg/l (analog); max. 0.05 to 10 mg/l (digital)
Measuring principle Membrane-covered, amperometric two-electrode system with integrated electronic components
Inflow speed Approx. 15 cm/s
Response time t90 Approx. 15 s
Operating temperature 0 to 45 °C
Settling time Chlorine dioxide: approx. 1 hr for initial startup; ozone: approx. 2 hrs for initial startup